2.Whilestillpinching,usetheotherhandtoplacetwongersabove the
collar on the applicator barrel so they are resting above the collar.
3. Place a thumb on the white plunger. Push the plunger down
completely,makingsureitisushagainsttheapplicatorbarrel.You
should hear 2 clicks. This inserts the needle and sensor under the
skin.
When you are pushing down on the plunger, do not pull back on
the collar.
4. Keep pinching up on the skin with one
hand. With the other hand, place two
ngersunder the collar. Keep the thumb
lightly on top of the white plunger, and
pull the collar back towards the thumb
until you hear 2 clicks or cannot pull
back any more. This leaves the sensor
under the skin and removes the needle
from the body.
5. Squeeze the center of the ribbed
release tabs on the sides of the sensor
pod to remove the applicator barrel.
Only the sensor pod will be left on the
body.
“2 CLICKS”
Steps 1-3. Push down the plunger - insert the needle and sensor
“ 2 CLICKS”
Step 4. Pull back the
collar - retract the needle
Step 5. Release the
applicator barrel