194 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
Hypoglycemia Alert Rate
The Alert Rate shows how often the alert is right or wrong. The True
AlertRateisthe%oftimethedevicealarmedwhenthebloodglucose
level was at or below the alert setting within 15 minutes before or after
thedevicealarmed.TheFalseAlertRateisthe%oftimethedevice
alarmed when the blood glucose level was above the alert setting within
15 minutes before or after the device alarmed.
For example, if you set the Low Glucose Alert to 70 mg/dL (3.9 mmol/L)
and your alarm sounds, how often can you expect your blood sugar to
actually be low? If your alarm sounds, you can expect your blood sugar
tobebelow70mg/dL(3.9mmol/L)approximately79%ofthetimeand
notbebelow70mg/dL(3.9mmol/L)approximately21%ofthetime
within the 15 minute period before or after your alarm sounds.
Hypoglycemia Detection Rate
The Detection Rate shows how often the device recognizes and alerts
you to an episode of hypoglycemia or how often it misses such an event.
TheHypoglycemiaDetectionRateisthe%oftimethebloodglucose
level was at or below the alert setting and device alarmed within 15
minutes before or after the blood glucose was at or below the alert
settings.TheHypoglycemiaMissedDetectionRateisthe%oftime
the blood glucose was at or below the alert setting, but the device did
not alarm within 15 minutes before or after the blood glucose was at or
below the alert setting.
For example, if you set the Low Glucose alert to 70 mg/dL (3.9 mmol/L),
how often will your alarm alert you if your blood glucose goes below
70 mg/dL (3.9 mmol/L)? If your blood sugar goes below 70 mg/dL (3.9
mmol/L),youcanexpectyouralarmtosound83%ofthetimeandnotto
soundapproximately17%oftimewithinthe15minuteperiodbeforeor
after your blood sugar goes below 70 mg/dL(3.9 mmol/L).
The High Glucose Alert
Estimates of how well the adjustable High Glucose Alert performs are
presented in Table 6.