Dexcom G4 PLATINUM Professional System User’s Guide | 197
Table 7. Percentage of System Readings within YSI
Values with Data Stratied in 2-Hour Increments after
Calibration
Time from
Calibration
Number
of paired
CGM-YSI
Percent
within
15/15%
YSI
Percent
within
20/20%
YSI
Percent
within
30/30%
YSI
Percent
within
40/40%
YSI
Percent
greater
than
40/40% YSI
0-2 hours 1929 78% 88% 96% 98% 2%
2-4 hours 1516 69% 81% 91% 96% 4%
4-6 hours 1547 69% 79% 91% 95% 5%
6-8 hours 1520 68% 79% 92% 97% 3%
8-10 hours 1555 71% 82% 92% 96% 4%
10-12 hours 1068 65% 77% 91% 96% 4%
12-14 hours 17 65% 76% 82% 88% 12%
NOTE: CGM readings are within 40 to 400 mg/dL (2.2-22.2 mmol/L),
inclusive.
Sensor Stability
Sensors can be worn for up to 7 days. To verify sensor performance
over time, 108 sensors were evaluated across the 7-day wear period.
Performance was estimated by calculating the percentage of System
readingswithin15mg/dL(0.9mmol/L)or15%,20mg/dL(1.1mmol/L)
or20%,30mg/dL(1.7mmol/L)or30%,40mg/dL(2.2mmol/L)or40%
andgreaterthan40mg/dL(2.2mmol/L)or40%oftheYSIvaluesat
the beginning (Day 1), middle (Day 4) and end (Day 7) of the System
lifecycle. The average and median of the absolute percent differences
are included in Table 8 showing consistent accuracy and sensor stability
over the 7-day life of the sensor.