196 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
event.TheHyperglycemiaDetectionRateisthe%oftimetheblood
glucose level was at or above the alert setting and the device alarmed
within 15 minutes before or after the blood glucose was at or above the
alertsettings.TheHyperglycemiaMissedDetectionRateisthe%of
time the blood glucose was at or above the alert setting, but the device
did not alarm within 15 minutes before or after the blood glucose was at
or above the alert setting.
For example, if you set your High Glucose alert to 200 mg/dL (11.1
mmol/L), how often will your alarm alert you if your blood glucose goes
at or above 200 mg/dL (11.1 mmol/L)? If your blood sugar goes above
200mg/dL(11.1mmol/L),youcanexpectyouralarmtosound97%of
thetimeandnottosoundapproximately3%oftimewithinthe15minute
period before or after your blood sugar goes above 200 mg/dL (11.1
mmol/L).
Calibration Stability
The System must be calibrated every 12 hours. To demonstrate
performance of the System over a 12-hour calibration period, sensors
were evaluated to verify that performance remains consistent over
the 12-hour calibration period. Systems were evaluated in 2-hour
increments after calibration. Performance was estimated at each 2-hour
intervalandstratiedbyglucoseconcentrationsbycalculatingthe
percentageofSystemreadingswithin15mg/dL(0.9mmol/L)or15%,
20mg/dL(1.1mmol/L)or20%,30mg/dL(1.7mmol/L)or30%,40mg/dL
(2.2mmol/L)or40%andgreaterthan40mg/dL(2.2mmol/L)or40%of
the YSI values in Table 7.