Dexcom G4 PLATINUM Professional System User’s Guide | 119
Regardlessofwhichalertproleyouchoose,ifyouconrmtherst
vibrate alert, you will not get a sound alert.
When you choose the prole setting this setting applies to all
alerts, alarm and prompts.
Withineachprolesetting,eachalerthasitsownuniquesoundpattern,
tone and volume level. This lets you to easily identify each alert and
alarm and its meaning.
Thexedlowalarmat55mg/dLcannotbeturnedofforadjusted.Itis
disabled in blinded mode.
Thesoft,normal,attentiveandHypoRepeatproleshavethefollowing
alert sequence:
•Therstalertisvibrateonly.
•Ifthealertisnotconrmedinveminutes,thesystemvibratesand
beeps.
•Ifthealertisnotconrmedinvemoreminutes,thesystemvibrates
andbeepslouder.Thiscontinuesatthesamevolumeeveryve
minutesuntilconrmed.
For the HypoRepeat prole in real-time mode only:
•Ifthealertisconrmedandthesensorglucosereadingscontinueto
be at or below 55 mg/dL the system repeats the alert sequence in 30
minutes.
Press the SELECTbuttontoconrmanyalert.
9.3.1 ALERT PROFILE OPTIONS
Vibrateprole:whenyouwanttosilencethereceiverandbe
alertedbyvibration.Theonlyexceptiontothisisthexed
low alarm at 55 mg/dL in real-time mode, which alerts you
asavibrationrst,followedbybeeps5minuteslaterifnot
conrmed.
(continued on next page)