120 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
9
The“TryIt”featureisfoundundertheProlesmenuandletsyouhear
an example of each alert and alarm.
9.3.2 ALERT PROFILE DETAILS
PROFILE
TYPE
VIBRATE SOFT NORMAL ATTENTIVE HYPOREPEAT
High Alert
In Real-Time
Mode Only
2 long
vibrates
2 long
vibrates
+ 2 low
beeps
2 long
vibrates +
2 medium
beeps
2 long
vibrates +
ascending
melody
2 long vibrates
+ 2 medium
beeps
(continued on next page)
“HypoRepeat”prole:verysimilartothenormalprole,but
itcontinuouslyrepeatsthexedlowalarmevery5seconds
until the sensor glucose reading rises above 55 mg/dL or
isconrmed.Thisproleisonlyapplicableifthereceiveris
used in real-time mode and can be helpful if you want extra
alerts for severe low sensor glucose readings.
ALERT PROFILE OPTIONS (continued from page
before)
Normalprole:thedefaultprolewhenyoureceivethe
system.Thisprolesetsallalertsandalarmstohigher
volume beeps.
Attentiveprole:whenyouneedthealerttobethemost
noticeable.Thisprolesetsallthealertsandalarmtoloud
and highly distinctive melodies.
Softprole:whenyouneedthealerttobelessnoticeable.
Thisprolesetsallthealertsandalarmtolowervolume
beeps.