Dexcom G4 Professional User's Guide Page 162

  • Download
  • Add to my manuals
  • Print
  • Page
    / 210
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 161
162 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System
or less than the SMBG values and the Median Absolute Percent
DifferenceshowsthathalfofthetimetheSystemreadsabout11.0%or
less different than SMBG values (Table 12).
Adverse Events
No serious adverse events or device-related serious adverse events
occurred during the study. Mild skin irritation, such as erythema or
edema, occurred in low frequency around adhesive area. No infection,
bruising, or bleeding occurred at the sensor needle insertion area or the
adhesive area.
14.2 Product Specications
Sensor Product Specications
Glucose Range 40 - 400 mg/dL
Sensor Life Up to 7 days
Calibration Commercially available blood glucose meter
Calibration Range 40 - 400 mg/dL
Storage
Condition
Temperature: 36° F - 77° F
Humidity:15%-85%RH
Sterilization Sterile by radiation
Transmitter Product Specications
Part Number P/N 9438-01
Dimensions
(including
sensor pod)
Length: 1.5 inches
Width: 0.9 inches
Thickness: 0.5 inches
Weight (including
sensor pod)
0.4 ounces
(continued on next page)