CHAPTER 14: TECHNICAL INFORMATION
14.1 DEVICE PERFORMANCE CHARACTERISTICS
NOTE: We recommend that you review the information in this chapter
with your healthcare provider to understand how well the Dexcom G4
PLATINUM System performs.
The Dexcom G4 PLATINUM System (the System) uses a glucose
sensor to continuously measure and monitor your glucose levels. The
sensor is “calibrated” using a commercially available blood glucose
meter; and once calibrated the System reports glucose readings up
to every 5 minutes. The System was evaluated in a clinical study in
which System readings were compared to blood glucose values to
assess its performance and how well the System readings compare
to a laboratory test method that measures blood glucose values.
Additionally, patients performed self-monitoring blood glucose
meter tests at home to assess the System performance in real use
environment.
Although the performance characteristics of the System are presented
in the following, there is no commonly accepted statistical approach for
capturing performance of continuous glucose monitors (CGMs), such
as the Dexcom G4 PLATINUM System.
Clinical Study Overview
The System performance was evaluated in a prospective clinical study;
72 participants were enrolled. All participants had Type 1 or Type 2
diabetes mellitus, and required insulin or oral medication to manage
theirdiabetes.About83%ofparticipantshadType1diabetesand
about17%ofparticipantshadType2diabetes.Thestudyincluded
subjects greater than 18 years of age.
Participants used the System for seven days. Thirty-six participants
each wore 2 sensors. Sensors were calibrated approximately once
every 12 hours, using the self-monitoring blood glucose (SMBG)
meter values obtained from the LifeScan® OneTouch® Ultra®2 meter.
Throughout the 7-day wear period, the sensor was calibrated with an