Dexcom, Inc. 6340 Sequence Drive San Diego, CA 92121 USA 1.858.200.0200 1.877.339.2664 www.dexcom.comLBL012528 Rev 03 MT22690CONTINUOUS GLUCOSE
8 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemGLOSSARY (continued from page before)Proles Sound pattern and volume level settings for
98 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1. Press the SELECT button to turn on the receiver. The 3-hour trend graph shows.2. Pr
Dexcom G4 PLATINUM User’s Guide | 991. From the alerts menu, press the UP or DOWN button to select “High Alert” or “Low Alert” and press the SELECT
100 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System95. To change this number, press the SELECT button, and then press the UP or DOWN butt
Dexcom G4 PLATINUM User’s Guide | 1019.2.1 SETTING A SNOOZE TIME FOR YOUR HIGH AND LOW GLUCOSE ALERTS The snooze feature lets you delay your high an
102 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemIf you set your fall rate to 2 mg/dL per minute and your sensor glucose readings fall a
Dexcom G4 PLATINUM User’s Guide | 1033. Press the UP or DOWN button to choose “On” or “Off.” Then, press the SELECT button to select “On” or “Off.”4
104 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9glucose readings. The out of range symbol in the upper right corner of the trend gra
Dexcom G4 PLATINUM User’s Guide | 1054. Press the UP or DOWN button to choose the amount of time out of range after which the receiver will alert. P
106 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9Thesoft,normal,attentiveandHypoRepeatproleshavethefollowingalert sequence:
Dexcom G4 PLATINUM User’s Guide | 107ALERT PROFILE OPTIONS (continued from page before)“HypoRepeat”prole:verysimilartothenormalprole,butit
Dexcom G4 PLATINUM User’s Guide | 9GLOSSARY (continued from page before)System ReadingA sensor glucose reading shown on your receiver. This reading i
108 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9ALERT PROFILE DETAILS (continued from page before)PROFILE TYPEVIBRATE SOFT NORMAL ATTE
Dexcom G4 PLATINUM User’s Guide | 1091. Press the SELECT button to turn on the receiver. The 3-hour trend graph will show.2. Press the SELECT butto
110 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 11110chapter tenEVENTS
112 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System10
Dexcom G4 PLATINUM User’s Guide | 113CHAPTER 10: EVENTSThe Events feature lets you record information that may help you and your healthcare professi
114 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System4. Press the UP or DOWN button to choose the event you want: “Carbs,” “Insulin,” “Exer
Dexcom G4 PLATINUM User’s Guide | 1154. Press the SELECTbuttontoconrmthe entry.10.1.3 CARBOHYDRATESThe Carbs event lets you enter the amount o
116 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System4. Press the LEFT or RIGHT button to choose either“OK”toconrmor“Cancel”todisca
Dexcom G4 PLATINUM User’s Guide | 1174. Press the LEFT or RIGHT button to choose either“OK”toconrmthisentryor“Cancel”to discard this entry,
10 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
118 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System3. Press the UP or DOWN button to enter your exercise duration (0-360 minutes), and pr
Dexcom G4 PLATINUM User’s Guide | 1192. Press the UP or DOWN button to choose your health event, and press the SELECT button.3. Check that the da
120 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 12111chapter elevenENDING A SENSOR SESSION
122 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System11
Dexcom G4 PLATINUM User’s Guide | 123CHAPTER 11: ENDING A SENSOR SESSIONYour sensor gives you sensor glucose readings for up to seven days. The perf
124 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System11Youcansetthesealertswiththeprolessetting(seeChapter9,Section9.3.2,Aler
Dexcom G4 PLATINUM User’s Guide | 125When you remove the sensor, make sure to pull out the sensor pod while the transmitter is still attached.1. Gen
126 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1111.3 TRANSMITTER REMOVALDo not remove the transmitter while the sensor pod is still
Dexcom G4 PLATINUM User’s Guide | 12712chapter twelveTAKING CARE OF YOURDEXCOM G4 PLATINUM SYSTEM
Dexcom G4 PLATINUM User’s Guide | 111chapter oneDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM DESCRIPTION
128 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System12
Dexcom G4 PLATINUM User’s Guide | 129CHAPTER 12: TAKING CARE OF YOUR DEXCOM G4 PLATINUM SYSTEM12.1 MAINTENANCETransmitter • Wipe the outside of t
130 | Dexcom G4 PLATINUM Continuous Glucose Monitoring Systemformat is YYYY-MM-DD. Insert sensors on or before the end of the calendar day printed o
Dexcom G4 PLATINUM User’s Guide | 13113chapter thirteenTROUBLESHOOTING
132 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13
Dexcom G4 PLATINUM User’s Guide | 133CHAPTER 13: TROUBLESHOOTINGThischapterprovideshelpfultipsandinstructionstoxissuesyoumayhave while
134 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13 • I am not able to remove the transmitter latch: - Hold the sensor pod with one
Dexcom G4 PLATINUM User’s Guide | 135 • See Chapter 7, Calibrating Your Dexcom G4 PLATINUM System, for further information.13.2.1 TYPES OF CALIBRA
136 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1313.2.2 WHAT TO DO FOR CALIBRATION PROMPTS1. When you see a calibration prompt, pres
Dexcom G4 PLATINUM User’s Guide | 137SELECT button to clear the screen, this symbol shows in the status area.If you see this screen, wait at least
12 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1
138 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13 • Make sure your sensor pod is not dislodged or peeling up. • Make sure your tran
Dexcom G4 PLATINUM User’s Guide | 139If the difference between your sensor glucose reading and blood glucose value is greater than 20% of the blood g
140 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13glucose to calibrate. Do not switch your meter in the middle of a sensor session. Blo
Dexcom G4 PLATINUM User’s Guide | 141 • Make sure your transmitter is snapped in. • Make sure your sensor pod is not dislodged or peeling up. •
142 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System132. Press the SELECT button with “OK” highlightedtoconrm.3. The Stop Sensor “thin
Dexcom G4 PLATINUM User’s Guide | 143This screen means you recently turned Share “On,” and your receiver is having trouble connecting with your iPhon
144 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System133. Charge your receiver.4. You may need to reset the time and date (see Chapter 4, S
Dexcom G4 PLATINUM User’s Guide | 14513.9.3 TRANSMITTER LOW BATTERYThis screen shows when the transmitter nears the end of its battery life (see Cha
146 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13This screen means your receiver and transmitter are not communicating and you are not
Dexcom G4 PLATINUM User’s Guide | 14713.11 ALERTS ARE NOT WORKING1. Make sure you have not turned off the sound and/or vibrations for the alerts. C
Dexcom G4 PLATINUM User’s Guide | 13CHAPTER 1: DEXCOM G4® PLATINUM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM DESCRIPTION1.1 SYSTEM CONTENTS: • sen
148 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 14914chapter fourteenTECHNICAL INFORMATION
150 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14
Dexcom G4 PLATINUM User’s Guide | 15116CHAPTER 14: TECHNICAL INFORMATION14.1 DEVICE PERFORMANCE CHARACTERISTICSNOTE: We recommend that you review t
152 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14and required insulin or oral medication to manage their diabetes. In the Original Stu
Dexcom G4 PLATINUM User’s Guide | 15316SW10586 SW10505 SW10050Includes Software 505 Incudes Software 505Does not include Software 505; includes Origi
154 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 1. System Agreement to YSI within CGM Glucose RangesCGM Glucose Range1 (mg/dL)
Dexcom G4 PLATINUM User’s Guide | 15516Agreement Relative to YSIAgreement between the System and blood glucose values is characterized using paired S
156 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14(SW10050) or Software 505 (SW10505).Agreement When CGM Reads “LOW” or “HIGH”The Syste
Dexcom G4 PLATINUM User’s Guide | 157System does not display glucose values below 40 mg/dL or above 400 mg/dL, the comparisons to the actual blood gl
14 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPRECAUTIONThe Dexcom G4 PLATINUM Sensor, Transmitter, and Receiver are not compatible wi
158 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 3-A. Concurrence of CGM Readings and YSI Values (Original Study)CGM (mg/dL)YSI
Dexcom G4 PLATINUM User’s Guide | 159Table 3-B. Concurrence of CGM Readings and YSI Values (Software 505 Study)CGM (mg/dL)YSI (mg/dL)Row percentage
160 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Another estimate used to show the accuracy of the System is the absolute percent diff
Dexcom G4 PLATINUM User’s Guide | 16116Table 4. System Difference to YSI within CGM Glucose RangesCGMGlucose Ranges1 (mg/dL)Study2 Number of Paired
162 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Low and High Glucose AlertsThe ability of the System to detect high and low glucose l
Dexcom G4 PLATINUM User’s Guide | 16316Table 5. Hypoglycemic Alert EvaluationHypoglycemicAlert Level(mg/dL)Study1True Alert RateFalse Alert RateHypo
164 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 6. Hyperglycemic Alert EvaluationHyperglycemicAlert Level(mg/dL)Study1True Ale
Dexcom G4 PLATINUM User’s Guide | 165Hyperglycemia Alert RateThe Alert Rate shows how often the alert is right or wrong. The True Alert Rate is the %
166 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14your alarm to sound 98% of the time and not to sound approximately 2% of time within
Dexcom G4 PLATINUM User’s Guide | 167Table 7. Percentage of System Readings within YSI Values with Data Stratied in 2-Hour Increments after Calibra
Dexcom G4 PLATINUM User’s Guide | 151.3 SENSOR OVERVIEWThe sensor is the piece that comes in a sterile, sealed sensor pouch. The sensor is made up o
168 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 8. Sensor Stability (Accuracy over Time)Day of Wear1Study2Number of paired CGM
Dexcom G4 PLATINUM User’s Guide | 169Precision of System ReadingsIn the Original Study, 36 subjects wore two Systems. This was to look at how similar
170 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemTable 9. Number of Readings Provided by Each Sensor Over 7-Days% of Total Possible Rea
Dexcom G4 PLATINUM User’s Guide | 17114Table 10. System Readings Within Wear DaysStatistic Study1Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7All Days2M
172 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 11. System Agreement to SMBG Within CGM Glucose RangesCGM Glucose Ranges1(mg/d
Dexcom G4 PLATINUM User’s Guide | 173Agreement and Accuracy Relative to SMBGDuring the study, agreement between the System and blood glucose values i
174 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Table 12. CGM System Difference to SMBG within CGM Glucose RangesCGM Glucose Ranges1
Dexcom G4 PLATINUM User’s Guide | 175Table 12 is categorized within CGM glucose ranges. Overall, the System in the Original Study reads, on average,
176 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1414.2 PRODUCT SPECIFICATIONSSensor Product SpecicationsGlucose Range 40 - 400 mg/dLS
Dexcom G4 PLATINUM User’s Guide | 177Moisture ProtectionIP28: temporary submersionProtection Against Electrical ShockType BF applied partTransmitter
16 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1.4 TRANSMITTER OVERVIEWThe transmitter is the gray, plastic “chip” that snaps into you
178 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14device can withstand exposure to common electrostatic (ESD) and electromagnetic inter
Dexcom G4 PLATINUM User’s Guide | 179Transmitter Electromagnetic Immunity Specications (continued from page before)Immunity Test IEC 60601 Test Leve
180 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Receiver Product SpecicationsPart Number MT22495Reading Frequency Every 5 minutesDim
Dexcom G4 PLATINUM User’s Guide | 181Guidance and Manufacturer’s Declaration – Electromagnetic ImmunityThe receiver (P/N MT22495) is intended for use
182 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Receiver Electromagnetic Immunity Specications(continued from page before)Immunity T
Dexcom G4 PLATINUM User’s Guide | 183System Electromagnetic Immunity SpecicationsImmunity Test IEC 60601 Test LevelReceiver Compliance LevelElectrom
184 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14System Electromagnetic Immunity Specications (continued from page before)Immunity Te
Dexcom G4 PLATINUM User’s Guide | 185a.Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio (cellular/cordless) telephones and l
186 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Electromagnetic Emissions Specications (continued from page before)Emissions Test Co
Dexcom G4 PLATINUM User’s Guide | 187Minimum Recommended Distance Between Transmitter and ReceiverRated MaximumOutput Power of Transmitter (W)Separa
Dexcom G4 PLATINUM User’s Guide | 171.5 RECEIVER OVERVIEWThe receiver is the small hand-held device that looks like a cell phone. It is about 4 inc
188 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14Power Supply/Charger SpecicationsDexcom P/N MT21255Class IIInput: AC Input 100-240
Dexcom G4 PLATINUM User’s Guide | 18914interference in a residential installation. This equipment generates uses and can radiate radio frequency ener
190 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System14PAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 19115chapter fteenUSER ASSISTANCE
192 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System15
Dexcom G4 PLATINUM User’s Guide | 19317CHAPTER 15: USER ASSISTANCE Dexcom Website: www.dexcom.com Dexcom Address: 6340 Sequence Drive San Diego,
194 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 19516chapter sixteenWARRANTY
196 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System16
Dexcom G4 PLATINUM User’s Guide | 197CHAPTER 16: WARRANTYDexcom G4 PLATINUM System Limited Warranty1. What is Covered and for How Long?Dexcom, Inc.
18 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemYour receiver and transmitter wirelessly pair together to communicate securely and only
198 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System16operation, maintenance, installation or adjustment; (f) installation, maintenance, an
Dexcom G4 PLATINUM User’s Guide | 199AND DISCLAIMS ALL SUCH OTHER WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTY OF MERCHANTABILITY, FITNESS
200 | Dexcom G4 PLATINUM Continuous Glucose Monitoring Systemthe original purchaser that the Dexcom G4 PLATINUM Transmitter is free from defects in
Dexcom G4 PLATINUM User’s Guide | 201of which was included with your Dexcom G4 PLATINUM CGM System and may be found at www.dexcom.com.3. What Are Dex
202 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPURPOSE, OR NON-INFRINGEMENT. EXCEPT TO THE EXTENT PROHIBITED BY APPLICABLE LAW, DEXCOM
Dexcom G4 PLATINUM User’s Guide | 20317chapter seventeenTRAVEL INFORMATION
204 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System17
Dexcom G4 PLATINUM User’s Guide | 205CHAPTER 17: TRAVEL INFORMATIONIt is safe for you to go through the metal detector or be “handwanded” while wear
206 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 20718chapter eighteenAPPENDIX
Dexcom G4 PLATINUM User’s Guide | 192chapter twoINDICATIONS FOR USE AND SAFETY STATEMENT
208 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18
Dexcom G4 PLATINUM User’s Guide | 209CHAPTER 18: APPENDIXAPPENDIX I, RECEIVER ALERTS, ALARM AND PROMPTSThe following tables describe the alarm, aler
210 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the
Dexcom G4 PLATINUM User’s Guide | 211Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al
212 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the
Dexcom G4 PLATINUM User’s Guide | 213Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al
214 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the
Dexcom G4 PLATINUM User’s Guide | 21520Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt,
216 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the
Dexcom G4 PLATINUM User’s Guide | 217Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al
20 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System2
218 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemReceiver Alerts, Alarm and Prompts (continued from page before)What will I see on the r
Dexcom G4 PLATINUM User’s Guide | 219Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al
220 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System18Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the
Dexcom G4 PLATINUM User’s Guide | 221Receiver Alerts, Alarm and Prompts (continued from page before)What will I see on the receiver screen?Prompt, al
222 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemAPPENDIX II, INDEXAAlarm, Low Glucose ...97Alerts ...
Dexcom G4 PLATINUM User’s Guide | 223C Calibration ...71 Calibration Update ...78 How
224 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemM Maintenance, System...129Menu Options...
Dexcom G4 PLATINUM User’s Guide | 225 Insertion Site ... 58 Overview ...15 Pod ...
226 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemLatch ... 9, 63 Overview ...
Dexcom G4 PLATINUM User’s Guide | 227APPENDIX III, SYMBOLS USED IN LABELINGThe following symbols may be found on the sensor, transmitter, and receive
Dexcom G4 PLATINUM User’s Guide | 21CHAPTER 2: INDICATIONS FOR USE AND SAFETY STATEMENT2.1 INDICATIONS FOR USEThe Dexcom G4 PLATINUM Continuous Glu
228 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemSymbols (continued from page before)ManufacturerEC REPAuthorized Representative in the
Dexcom, Inc. 6340 Sequence Drive San Diego, CA 92121 USA 1.858.200.0200 1.877.339.2664 www.dexcom.comLBL012528 Rev 03 MT22690CONTINUOUS GLUCOSE
22 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System22.3 CONTRAINDICATIONS• Remove the Dexcom G4 PLATINUM Sensor, Transmitter, and Receive
Dexcom G4 PLATINUM User’s Guide | 23values may differ from sensor glucose readings. Using the sensor glucose readings for treatment decisions could l
24 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System2• Sensor placement and insertion is not approved for sites other than the belly (abdom
Dexcom G4 PLATINUM User’s Guide | 25to measure glucose. Insertion in those areas might affect sensor performance, and you might miss a low or high bl
26 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System2ischangingatasignicantrate(e.g.,2-3mg/dL/minormore than 3 mg/dL each minute)
Dexcom G4 PLATINUM User’s Guide | 27• The Dexcom G4 PLATINUM Sensor, Transmitter, and Receiver are not compatible with the SEVEN/SEVEN PLUS Transmit
Dexcom G4 PLATINUM User’s Guide | 1USER’S GUIDEDEXCOM G4® PLATINUMCONTINUOUS GLUCOSE MONITORING SYSTEMRECEIVER WITH ShareTMIMPORTANT CONTACTS AND NU
28 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 293chapter threeRISKS AND BENEFITS
30 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System13
Dexcom G4 PLATINUM User’s Guide | 31CHAPTER 3: RISKS AND BENEFITS3.1 RISKSThere are some known risks with using real-time CGM. You will not get sen
32 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System3events happen, you might feel discomfort in the area the sensor is inserted.There is a
Dexcom G4 PLATINUM User’s Guide | 33A1c values, reducing glycemic variability and time spent in low and high blood glucose ranges)1, 2 , 3 which can
34 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 354chapter fourCHARGING YOUR RECEIVER AND THE RECEIVER MAIN MENU
36 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System4
Dexcom G4 PLATINUM User’s Guide | 37CHAPTER 4: CHARGING YOUR RECEIVER AND THE RECEIVER MAIN MENU4.1 CHARGING YOUR RECEIVER BATTERYThe receiver batt
2 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEMtable of contentsGLOSSARY ...
38 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemCharge your receiver battery before each sensor insertion. Periodically check your batte
Dexcom G4 PLATINUM User’s Guide | 395. The battery charging screen will show on the receiver.6. After a few seconds the trend graph will show with
40 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System44. After a few seconds, the trend graph screen will show the battery charging symbol i
Dexcom G4 PLATINUM User’s Guide | 412Receiver Main Menu Options (continued from page before)Menu PurposeUser’s Guide ReferenceStart SensorTo start a
42 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 435chapter veDEXCOM G4 PLATINUM SYSTEM SETUP
44 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System5
Dexcom G4 PLATINUM User’s Guide | 45CHAPTER 5: DEXCOM G4 PLATINUM SYSTEM SETUPThischapterhelpsyouwhenyourstsetupyourContinuousGlucoseMo
46 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System5 (3) Press the SELECT button after you enter the last number or lettertoconrm
Dexcom G4 PLATINUM User’s Guide | 475.2 THE SETTINGS MENUThe Settings menu lets you change the time, date or transmitter ID. The SetupWizardonlyw
Dexcom G4 PLATINUM User’s Guide | 3CHAPTER 5: DEXCOM G4 PLATINUM SYSTEM SETUP 435.1 Setting Up the Receiver and Pairing with Your Transmitter ...
48 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System52. Press the RIGHT or SELECT button to highlight each value in the date and time.3.
Dexcom G4 PLATINUM User’s Guide | 491. From the Settings menu, press the UP or DOWN button to scroll to “Transmitter ID” and press the SELECT button
50 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System55.3 CHECKING INFORMATION ABOUT YOUR DEXCOM G4 PLATINUM SYSTEMYou can check your receiv
Dexcom G4 PLATINUM User’s Guide | 51PRECAUTIONThe transmission range from the transmitter to the receiver is up to 20 feet without obstruction. Wirel
52 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System5Press the SELECT, LEFT or RIGHT button to see the trend graph. This antenna symbol show
Dexcom G4 PLATINUM User’s Guide | 536chapter sixINSERTING A SENSOR AND STARTING A SENSOR SESSION
54 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6
Dexcom G4 PLATINUM User’s Guide | 55CHAPTER 6: INSERTING A SENSOR AND STARTING A SENSOR SESSIONYou need a sensor, a transmitter, and a receiver to u
56 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6.1 BEFORE YOU START• Make sure the correct transmitter ID has been entered into your
Dexcom G4 PLATINUM User’s Guide | 576.2 REMOVING THE SENSOR FROM ITS PACKAGING• Wash your hands thoroughly, and dry them.• Carefully remove the sen
4 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8.3 Glucose Status Area Symbols ...89CHAPTER 9: ALERTS, ALARM
58 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System66.3 CHOOSING AN INSERTION SITEReview the tutorial disc for more help to learn how to i
Dexcom G4 PLATINUM User’s Guide | 596.4 PLACING THE SENSOR1. Clean your skin at the sensor placement site with an alcohol wipe. Make sure the area
60 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System63. Place the sensor horizontally, NOT vertically, on your skin. 4.Moveyourngersa
Dexcom G4 PLATINUM User’s Guide | 612.Whilestillpinching,useyourotherhandtoplacetwongersabove the collar on the applicator barrel so th
62 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6 • Make sure the transmitter latch is down (against your body) before squeezing the t
Dexcom G4 PLATINUM User’s Guide | 63 c. With your other hand, pull the transmitter latch up and forward, over the transmitter, to snap it into plac
64 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System63. Press the DOWN button to highlight “Start Sensor.” • The “Start Sensor” menu opti
Dexcom G4 PLATINUM User’s Guide | 656.8 SENSOR STARTUP PERIODThe sensor needs a 2-hour startup period to adjust to being under the skin.Your trend g
66 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System6time to calibrate your sensor. Chapter 7, Calibrating Your Dexcom G4 PLATINUM System, t
Dexcom G4 PLATINUM User’s Guide | 672.Conrmthatyouwanttoshutdownyour receiver. a. If you want to shut down, press the LEFT button to high
Dexcom G4 PLATINUM User’s Guide | 512.2 Storage . ... 12912.3 Product Disposal . ...
68 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 697chapter sevenCALIBRATING YOUR DEXCOM G4 PLATINUM SYSTEM
70 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7
Dexcom G4 PLATINUM User’s Guide | 71CHAPTER 7: CALIBRATING YOUR DEXCOM G4 PLATINUM SYSTEMYou must calibrate Dexcom G4 PLATINUM System sensor glucose
72 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7Example Minimum Calibration Schedule During Seven-Day Sensor SessionPRECAUTIONDo not us
Dexcom G4 PLATINUM User’s Guide | 737.2 HOW TO CALIBRATEYou must enter the exact blood glucose value from your meter for each calibration. Blood glu
74 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7Only use blood glucose values between 40-400 mg/dL for calibration. If the blood glucos
Dexcom G4 PLATINUM User’s Guide | 751. Wash and dry your hands, make sure your glucose test strips are not expired and have been stored properly, an
76 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System77. Conrm that the blood glucose value you entered is correct. Entering incorrect valu
Dexcom G4 PLATINUM User’s Guide | 777.3 STARTUP CALIBRATIONTwo hours after you start the sensor session (see Chapter 6, Section 6.7, Starting a Sens
6 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemGLOSSARYAlternative Site BG TestingThis is when you take a blood glucose value on your me
78 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System7.4 12-HOUR CALIBRATION UPDATE Calibrateyoursystematleastevery12hoursafteryour
Dexcom G4 PLATINUM User’s Guide | 79When you see this calibration prompt it means it is time to calibrate with a single blood glucose value.Take 1 bl
80 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemPAGE INTENTIONALLY LEFT BLANK
Dexcom G4 PLATINUM User’s Guide | 818chapter eightSENSOR GLUCOSE READINGS AND TRENDS
82 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System1038
Dexcom G4 PLATINUM User’s Guide | 83CHAPTER 8: SENSOR GLUCOSE READINGS AND TRENDS This chapter teaches you how to view your sensor glucose readings
84 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8WARNINGDo not use the Dexcom G4 PLATINUM System for treatment decisions, such as how mu
Dexcom G4 PLATINUM User’s Guide | 85 • Each “dot” on the trend graph is a sensor glucose reading reported every 5 minutes. • The trend graph shows
86 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8Which Trend Graph Do You See?(Scroll up from the 3-hour graph to reach the 1-hour graph
Dexcom G4 PLATINUM User’s Guide | 87The receiver displays “LOW” when the most recent sensor glucose reading is less than 40 mg/dL and “HIGH” when the
Dexcom G4 PLATINUM User’s Guide | 7GLOSSARY (continued from page before)Dexcom Share™ SystemSecondary notication system to the Dexcom G4 PLATINUM Sy
88 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8Trend Arrows (continued from page before)Slowly rising: Your glucose is rising 1-2 mg/d
Dexcom G4 PLATINUM User’s Guide | 89top of the trend graph, the trend arrows will not show. If the trend arrow is missing, but you are concerned that
90 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System8Status Symbols (continued from page before)Glucose Reading ErrorThis symbol means the r
Dexcom G4 PLATINUM User’s Guide | 919chapter nineALERTS, ALARM & PROFILES
92 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9
Dexcom G4 PLATINUM User’s Guide | 93CHAPTER 9: ALERTS, ALARM & PROFILESThis chapter teaches you about your Dexcom G4 PLATINUM CGM System’s many
94 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System9Default Alert/Alarm Settings (continued from page before)What will I see on the receive
Dexcom G4 PLATINUM User’s Guide | 95Default Alert/Alarm Settings (continued from page before)What will I see on the receiver screen?What does this me
96 | Dexcom G4 PLATINUM Continuous Glucose Monitoring SystemWhen you have both your high and low alerts turned on, a gray zone on your trend graph s
Dexcom G4 PLATINUM User’s Guide | 97the alert (to change your snooze settings see Chapter 9, Section 9.2, Advanced Alerts).9.1.2.3 LOW GLUCOSE ALARM
Comments to this Manuals